OUTPUTS AFTER POWER ON OR RESET
PRODUCT LIST > i100 >
The analog/digital output states upon instruNet reset or power on are
described as follows. For information on how to deal with unexpected power outages
and computer crashes, please see Detecting Power Cycling or Computer Crash.
i200 Controller Card
i200 Digital Inputs
These start as inputs upon Reset. They are pulled high via a 10K internal resistor.
i200 Digital Outputs
These are driven by the CONTROLLER DIGITAL IN upon Reset, which are pulled high
via a 10K. So if nothing is tied to the corresponding CONTROLLER DIGITAL IN, these
will start as 1's. If you want these to power-up as 0's, send the CONTROLLER
DIGITAL OUT through an inverter (e.g. 74LS04).
i100 Data Acquistion Box
i100 Digital I/O
These are set to inputs upon power-up or reset, and are pulled high via a 10K resistor.
If they are wired as outputs, then they will appear as 1's upon power-up or reset. If you
want these to be outputs that power-up as 0's, send the CONTROLLER DIGITAL OUT
through an inverter (e.g. 74LS04).
i100 Analog Outputs
When the i100 powers off, these drop to 0V. Upon i100 power on, they stay at 0V, yet
when the i100 is auto-calibrated or reset by the host computer, they flicker to various
levels for approximately 1sec and then settle at 0V.
|